SPECIFICATIONS
Power supply
Display mode
Measurement mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the arm
Weight
External dimensions
Attachment
Mode of operation
Degree of protection
Protection against
ingress of water
Software version
Device classification
WARNING: No modification of this equipment is allowed.
ENGLISH
3.7V 1000mAH Built-in rechargeable
li-polymer battery, 6V
Digital LCD display
V.A.= 86.1mm(L) x24mm(W)
Oscillographic testing mode
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Pressure:
5°C-40°C within ±3mmHg (0.4kPa)
Pulse value: ±5%
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour partial
pressure greater than 50 hPa
An atmospheric pressure range of : 700 hPa to 1060 hPa
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
About 22cm-42cm
Approx.284g
Approx.130.9mm×73mm×29.4mm
AC Adaptor, user manual
Continuous operation
Type BF applied part
IP22: The first number 2: Protected against solid foreign
objects of 12,5mm Ф and greater. The second number:
Protected against vertically falling water drops when
enclosure titled up to 15º. Vertically falling drops shall
have no harmful effects when the enclosure is titled at any
angle up to 15º on either side of the vertical
A01
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor charged Mode:
Class II ME Equipment
1A AC Adaptor
52