Guidance and manufacturer's declaration – Immunity Emissions
The HOSPI PLUS SUCTION ASPIRATOR is intended for use in the electromagnetic environment specified below. The
customers or the user of the HOSPI PLUS SUCTION ASPIRATOR should assure that it's used in such an environment.
Immunity Test
Level indicated by
the EN 60601-1-2
Conducted
da 3Vrms 150kHz
Immunity
a 80MHz
EN 61000-4-6
(for non life-supporting
devices)
da 3V/m 80MHz
Radiated
a 2.7GHz
Immunity
(for non life-supporting
EN 61000-4-3
devices )
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied.
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by reflection from buildings, objects and people.
a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and ter-
restrial mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically
and accurately foreseen. To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-
magnetic study of the site should be considered. If the field intensity measured in the place where the device will be used
surpasses the above mentioned applicable level of conformity, the normal functioning of the device should be monitored.
If abnormal performance arises, additional measures such as changing the device's direction or positioning may be
necessary.
b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V/m.
ENGLISH
Livello di
conformità
V1 = 3 V rms
E1 = 10 V / m
Electromagnetic environments - guidance
Portable and mobile RF communication equipment, in-
cluding cables, should be used no closer to any part of
the device, than the recommended separation distance
calculated from the equation applicable to the frequen-
cy of the transmitter.
Recommended separation distance
3,5
√P
d=
V
1
from 80MHz to 800MHz
12
√P
d=
E
1
23
√P
from 800MHz to 2,7GHz
d=
E
1
Where P is the maximum output power rating of the
transmitter in Watt (W) according to the transmitter
manufacturer and is the recommended separation
distance in metres (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site
study of the site, should be less that the compliance
level in each frequency range. Interference
may occur in the vicinity of equipment
marked with the following symbol:
24