General Specifications
Display: 3-3/4 digit LCD, 4000
counts, 9999 counts in Frequency
mode, 40 segment bargraph with
function and unit annunciators.
Polarity Indication: Automatic,
positive implied, negative (–)
indicated
Zero Adjustment: Automatic
Overrange Indication: Most significant
digit (MSD) blinks. Other digits
shown as zero.
Low Battery Indication:
battery immediately.
Display Update Rate: 2/sec, nominal;
1/sec in capacitance and frequency
modes; 20/sec for bargraph.
Operating Temp.: 0 to 50°C, 0 to 70%
Relative Humidity
Storage Temp: -20°C to 60°C, 0 to
80% RH, battery removed
Temperature coefficient: 0.1 x (spec.
accuracy)/°C (<18°C or >28°C°
Power: Standard 9-volt battery, NEDA
1604, JIS 006P, IEC 6F22
Auto Power-Off: Meter turns off after
30 minutes of inactivity.
Battery Life (Typical): 500 hours
alkaline, 250 hours carbon-zinc
Dimensions (H x W x D): 7.5" x 3.5" x
1.3", (175 x 90 x 34 mm)
Weight (incl. battery): 568 g (20 oz)
Accessories: Test leads, spare fuse,
battery, and Operator's Manual
Safety: Meets EN61010-1 Cat III
1000V; EMC: Meets EN50081-1,
EN50082-2.
EMC:
complies with require-
ments of the following European
Community
89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC
SPECIFICATIONS
. Change
This
product
Directives:
– 28 –
(Low Voltage) as amended by
93/68/EEC (CE Marking).
However, electrical noise or
intense electromagnetic fields in
the vicinity of the equipment may
disturb the measurement circuit.
Measuring instruments will also
respond to unwanted signals that
may be present within the
measurement circuit. Users
should exercise care and take
appropriate precautions to avoid
misleading results when making
measurements in the presence of
electronic interference.
Electrical Specifications
Accuracies at 23°C ± 5°C, <75% RH.
DC Volts
Ranges: 400mV, 4, 40, 400, 1000V
Resolution, 100µV in 400mV range:
Accuracy, all Rgs: ±(0.25%rdg +2dgt)
Input Impedance, 400mV rng.
>1000 MW ; 4V rng: 11 MW ; 40V-
1000V rngs: 10 M W
OL Protection: 1100VDC or Peak AC.
6kV transient protection.
AC Volts (50-1000Hz).
Model 2015: True RMS meter. Accuracies
apply to all waveforms.
Ranges: 400mV, 4, 40, 400, 750V
Resolution: 100µV in 400mV range
Crest Factor 3:1
Accuracy ±(%rdg +nr digits):
Range
50-500Hz
400mV*
±(1.0% +5) unspecified
40V-400V ±(0.8% +3) ±(1.0% +5)
All other
±(0.8% +3) ±(1.2% +5)
Input Impedance,
400mV rg. >1000 MW ; 4V rg:
11 MW ; 40V-750V rgs: 10 M W
OL Protection: 1100VDC or Peak AC.
6kV transient protection.
500-1000Hz