Function Mode,
Continued
Voltage
Power Source
Battery Life
MR Mode
Modulated Rate: The pulse rate is automatically varied in
an interval of 0.5 seconds. The modulation range of pulse
rate is from setting value to 50% less than the control
setting value, then returns to the setting value. Pulse rate is
adjustable but pulse width is fixed.
B (Burst
Bursts consist of pulse width (adjustable),
Mode)
frequency=100Hz. Bursts occur twice every second.
©
Pulse width and pulse rate are adjustable.
(Constant
Mode)
SD Mode
Strength duration and width / rate modulation are
adjustable. SD Mode consists of alternating modulated
pulse width and pulse rate. One parameter is increasing
while the other parameter is decreasing and vice-versa. The
variety value is 50%; full cycle time is 22 seconds.
Alternate
Stimulation output will begin from channel 2, last for 12
Mode
seconds, then switch to channel 1 for 12 seconds. During
this mode, the stimulation will continue to alternate
between the two channels.
0 – 100 volt (open circuit)
9-volt battery (alkaline or nickel-cadmium rechargeable)
Approximately 70 hours at nominal settings