change to display label "STAT" prompting the current NIBP mode, therefore the MAP value will not be
displayed. When the "STAT" mode (at most lasting 5 minuter) finishes (or measurement error occurs or
it is interrupted manually), then the device will shift into "Manual" mode automatically.
² "NIBP Check ": press it to enter into NIBP verification setup screen, as shown in figure 4.13.
Note: password need to input to perform NIBP Check. Default password is "1234".
Screen description:
² "Verification mode 1": Pressure source is generated by internal pump. Move the cursor to NIBP
Verification mode 1 "Start" button, press the OK button to begin the pressure meter verification.
(Meanwhile, the "Start" shifts to "Stop", after the verification the "Stop" shifts to "Start").
² "Verification mode 2": Pressure source is coming from outside. Move the cursor to NIBP Verification
mode 2"Start" button, press the OK key to begin the pressure meter verification. (Meanwhile, the
"Start" shifts to "Stop", after the verification the "Stop" shifts to "Start").
² "Air leakage": Check the air leakage in the pneumatic system. Move the cursor to Air Leakage "Start"
button, press the OK key, the pump inflates to certain pressure and then the valve will be closed to
detect leakage for 10 seconds, then the pressure will be released automatically and the screen displays
the checking result
² If the following message pops up, then NIBP measurement should be stopped.
1) Pressure verification...
2) Air leakage preparing...
3) Air leakage countdown...
4) Air leakage in 10s:...
² "Hemostat": press it to enter into Hemostat setup screen, as shown in figure 4.14.
User Manual for Vital Signs Monitor
Figure 4.13 NIBP verification setup screen
39