Safety information
Ask a qualified technician to verify device
compatibility. Attaching any other device to the
Ethernet port may damage the incubator and may
not be safe for the user.
•Do not attempt to repair the Auto-reader by
yourself, that could lead to major and irreversible
damages to the device. In case of device
malfunction, contact your local distributor for
further assistance.
To reduce the risk of using incompletely
sterilized loads:
•Please
read,
understand,
Instructions for Use of each SCBI before its
incubation.
•Do not remove the SCBI before the incubator
reports the final readout result. Check a result
ticket is printed.
•Check that culture medium completely wets the
spore carrier.
To avoid the risk of injury, related to glass
fragments produced when crushing the glass
ampoule inside the SCBI tube:
•Cool the SCBI during the indicated time before
crushing the ampoule.
•Do not handle the SCBI excessively since this
might cause the glass ampoule to burst.
•Wear safety gloves and glasses when removing
the SCBI from the sterilizer, pressing the SCBI's
cap, and crushing the SCBI's ampoule.
•Do not use your finger for crushing SCBI. Use the
Ampoule Crusher instead.
To avoid a potentially hazardous situation:
•Avoid contact with the hot metal block inside
each incubation position.
•Do not insert your fingers, or any other element,
inside the incubation positions.
•Place only compatible indicators inside the
incubation positions.
To avoid SCBIs from absorbing fluorescent
particles:
•Avoid direct contact between the SCBIs and
Chemical Indicators or Tapes before the SCBI
incubation.
Important
Do not use this product in a manner not specified
by Terragene
S.A., otherwise the protection
®
provided by the product might be affected.
Only Terragene
S.A. authorized personnel can
®
access or service the internal components of
the Incubator. Parts or components inside the
Incubator should not be manipulated by the user.
Power supply specifications
Input
parameters
Voltage range
Frequency
*Current
Output parameters
Voltage
**Current
and
follow
the
* 1 Ampere in USA and Canada.
** 3 Amperes in USA and Canada.
Terragene
stabilizers, since they fulfill two functions: to stabilize and
maintain the energy during a power outage.
Environment operating conditions
Environmental
Conditions
Altitude
Operation temperature
Relative humidity
Installation/overvoltage
Degree of contamination
Storage temperature
Voltage
Bionova
following standards and directives:
Electrical Safety.
Electromagnetic
Compatibility (EMC).
European Commission.
Bionova
photobiological risk and do not generate dangerous optical
radiation in any of its normal operation conditions as per the
requirements of IEC 62471 Standard.
Designed under Quality Management System standards ISO
13485:2016/NS-EN ISO 13485:2016.
IC10/20FRLCD Page 6
Operating conditions
Operating
Condittion
(100-240)
50/60
0,2
Values
12
4
S.A. recommends the use of UPS instead of voltage
®
Operating
Conditions
3500 (máx.)
10-30
30-80
category II
2
10-30
12
Device Safety Compliance
IC10/20FRLCD Auto-reader complies with the
®
IEC 61010-1
IEC 61010-2-010
Low Voltage Directive
2014/35/EU.
EN 61326-1
EMC Directive 2014/30/EU
RoHS Directive 2011/65/EU.
WEEE Directive 2012/19/EU.
IC10/20FRLCD Auto-reader do not represent
®
Units
Volts
Hertz
Ampers
Units
DC Volts
Amperes
Units.
Meters
Celsius
%
Celcius
DC volts