Operation (cont'd)
Calibrate Resistance Inputs
Select the Source mode by pressing the SOURCE/READ button until "SOURCE" appears on the
display.
Display: Ω
This function simulates a resistance into a variety of instruments.
1.
Turn the Selector knob to OHMS.
2.
Connect the black and red leads to the meter as shown.
3.
Disconnect one or both input wires from the device to be calibrated.
4.
Connect the red lead of the meter to the plus (+) input of the device and the black lead to
the minus (-).
5.
With the toggle switch at SET, adjust the output to the desired value.
6.
Flip the toggle switch to HI or LO to use the stored values. See the Storing Output Values
section to change the stored values.
Receiver Input
Ohms or 2-Wire RTD
Controller
Transmitter
Computer