Mechanical
Size:
Weight:
Power Requirements
Internally Powered Equipment
Mode of Operation is Continuous
Rated Input:
Supply Frequency:
Supply Voltage:
Battery Operation
Type:
Operating Time:
Charge Time:
Low Battery Alert:
Safety
The ICP EXPRESS Monitor is Class I, CF Type Medical Equipment.
The input of ICP EXPRESS Monitor is protected against defibrillator discharge with no
interruption to normal functionality.
The ICP EXPRESS Monitor complies with:
IEC 60601-1 (2005) Medical Equipment, Part 1: General Requirements for Safety
IEC 60601-1-2 (2001) Electromagnetic Compatibility (EMC) Requirements and Tests

Cleaning and Maintenance

The ICP EXPRESS Monitor and ICP EXPRESS Cable can be cleaned using a clean
cloth dampened with 70% isopropyl alcohol or mild hospital cleaning detergent/
bactericide.
It is not necessary to perform any yearly maintenance or calibration of the ICP
EXPRESS unit.
CAUTION: DO NOT immerse the ICP EXPRESS Monitor or ICP EXPRESS Cable in any
liquid, or allow liquid to enter the plug or connections Do not use cable if connectors
become wet Erratic results can occur
CAUTION: Always disconnect the ICP EXPRESS Monitor from its power supply
before cleaning
Sterilization
CAUTION: DO NOT STERILIZE the ICP EXPRESS Monitor
CAUTION: DO NOT AUTOCLAVE the ICP EXPRESS Cable Sterilize the ICP EXPRESS
Cable using ethylene oxide (EtO) gas only.
The ICP EXPRESS Cable is compatible with ethylene oxide (EtO) gas sterilization only.
Suggested ethylene oxide sterilization parameters are as follows:
Temperature:
130°F (54.4°C)
Exposure Time:
2 hrs.
Aeration:
120°F (48.8°C) for 8 hrs
Always refer to the sterilizer manufacturer's instructions for correct times,
temperatures and pressure settings.
Service and Repair
For service, repair, or replacement outside the United States, contact your local sales
representative.
For service or repairs to the ICP EXPRESS Monitor within the United States, contact
your local sales representative directly.
6.1 in. H x 3.5 in. W x 6.2 in. D
4.75 lbs (2.15 kg)
120/127VAC, 0.2A
60 Hz
120/127 VAC
Lead Acid, rechargeable
3 hours, nominal (fully charged, new battery)
<12 hours (for fully discharged battery)
<15 minutes life remaining
End of Useful Life
The ICP Express Monitor contains electrical components. Consider recycling at the
end of service life. The product can be returned to Integra for recycling. Dispose of
the equipment in accordance with local ordinances.
Accessories
Use only Codman Microsensor ICP Transducers for intracranial pressure measurements.
The Codman Microsensor is available in kits tailored for different applications including
subdural, intraparenchymal and intraventricular ICP patient monitoring. Please contact
your local sales representative
The following ICP patient monitoring accessories are available:
Description
Codman Microsensor Basic Kit
Codman Microsensor Ventricular
Catheter Kit - with Tuohy-Borst Adapter
Codman Microsensor Ventricular
Catheter Kit
Codman ICP Express Cable
Codman Microsensor Metal Skull Bolt Kit
Codman Microsensor Metal Skull Bolt Kit
Codman CereLink
ICP Sensor Basic Kit
Codman CereLink ICP Sensor Metal
Skull Bolt Kit
Codman CereLink ICP Sensor Plastic
Skull Bolt Kit
Codman CereLink ICP Sensor
Ventricular Catheter Kit
PRODUCT INFORMATION DISCLOSURE
INTEGRA LIFESCIENCES CORPORATION ("INTEGRA") HAS EXERCISED REASONABLE CARE IN
THE SELECTION OF MATERIALS AND THE MANUFACTURE OF THESE PRODUCTS. INTEGRA
WARRANTS THAT THESE PRODUCTS SHALL CONFORM TO THE PRODUCT LIMITED WARRANTY
AS PROVIDED IN THE PRODUCT LABELING OR APPLICABLE PRODUCT CATALOG. THIS WARRANTY
IS EXCLUSIVE, AND INTEGRA DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESSED OR
IMPLIED, INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE. INTEGRA SHALL NOT BE LIABLE FOR ANY INCIDENTAL
OR CONSEQUENTIAL LOSS, DAMAGE, OR EXPENSE, DIRECTLY OR INDIRECTLY ARISING FROM USE
OF THESE PRODUCTS. INTEGRA NEITHER ASSUMES NOR AUTHORIZES ANY PERSON TO ASSUME
ANY OTHER OR ADDITIONAL LIABILITY OR RESPONSIBILITY IN CONNECTION WITH THESE
PRODUCTS.
8
EN – ENGLISH
Codman Catalog No
82-6631, 626631, 626631US
82-6633, 626633, 626633US
82-6653, 626653, 626653US
82-6636
82-6638, 626638, 626638US
82-6639 (5 per case)
82-6850
82-6851
82-6852
82-6854

Publicidad