Electromagnetic Compatibility - DTF PURENEB AERODJINN Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27

13. Electromagnetic compatibility

Electro medical devices require special precautions against EMC and must be installed and set to operate in
accordance with the EMC information provided in this manual.
WARNING: Avoid using this unit near or stacked onto other equipment as this may cause a malfunction. If this is
necessary, this appliance and the other appliances should be monitored for normal operation.
WARNING: Use of accessories, transducers and cables other than those specified or supplied by the manufacturer
of this device may cause an increase in electromagnetic emissions or a decrease in the immunity of this device
and cause improper operation.
WARNING: Portable RF communication devices (including peripherals such as antenna cables and external
antennas) should not be used closer than 12 inches (30 cm) to any part of the AERODJINN/AERODJINN+, including
cables specified by the manufacturer. Otherwise, performance may be impaired.
WARNING: If the device malfunctions due to electromagnetic disturbances, the performance of the device may be
impaired and result in delayed patient care.
Wireless communication devices such as wireless home networking devices, mobile phones, cordless phones and
their bases, walkie-talkies can affect the device, and it is recommended to keep them at a distance of at least 0.5
m from the device.
Emissions test
RF emissions (CISPR 11)
RF emissions (CISPR 11)
Harmonic emissions (IEC 61000-3-2)
Voltage fluctuations / flicker emissions (IEC 61000-3-3)
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
Voltage interruptions
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
Conducted RF
IEC 61000-4-6
Guidance and manufacturer's declaration – electromagnetic emissions
Guidance and manufacturer's declaration – electromagnetic immunity
Test level (IEC 60601)
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0% U
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% U
AND 70% U
Single phase : at 0°
0% U
: 250/300 cycles
T
80 MHz to 2,7 GHz
385 MHz at 27 V/m
450 MHz at 28 V/m
710, 745, 780 MHz at 9 V/m
810, 870, 930 MHz at 28 V/m
1720, 1845, 1970 MHz at 28 V/m
2450 MHz at 28 V/m
5240, 5500, 5785 MHz at 9 V/m
0,15 MHz – 80 MHz
6V in ISM bands and amateur radio bands
between 0.15 MHz and 80 MHz
80% MA at 1kHz
; 0,5 cycle
T
; 1 cycle
T
; 25/30 cycles
T
30 A/m
10 V/m
3V
18
Compliance
Group 1
Class B
Class A
Complies
Compliance level
± 8 kV
± 15 kV
± 2 kV
± 1 kV
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0% U
; 0,5 cycle
T
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% U
; 1 cycle
T
AND 70% U
; 25/30 cycles
T
Single phase : at 0°
0% U
: 250/300 cycles
T
30 A/m
10 V/m
80 MHz to 2,7 GHz
385 MHz at 27 V/m
450 MHz at 28 V/m
710, 745, 780 MHz at 9 V/m
810, 870, 930 MHz at 28 V/m
1720, 1845, 1970 MHz at 28 V/m
2450 MHz at 28 V/m
5240, 5500, 5785 MHz at 9 V/m
3V
0,15 MHz – 80 MHz
6V in ISM bands and amateur radio bands
between 0.15 MHz and 80 MHz