R R e e m m o o v v i i n n g g a a n n d d r r e e p p l l a a c c i i n n g g t t h h e e r r e e s s e e r r v v o o i i r r
The removable reservoir enables you to fill or drain the reservoir away from the controller without interrupting therapy. You
will need to have the reservoir installed before starting a therapy.
C C A A U U T T I I O O N N - - Do not place your fingers in between the reservoir and the sides of the controller, to avoid the risk of pinching
your fingers.
To remove the reservoir, pull forward at an angle, and lift out the reservoir (Figure 16).
A
B
C
F F i i g g u u r r e e 1 1 6 6 – – R R e e m m o o v v a a b b l l e e r r e e s s e e r r v v o o i i r r
1. To replace the reservoir, align the base of the reservoir over the drain (C).
2. Align the notch on the back of the reservoir (A) with the hook on the controller (B) (Figure 16)
3. Push the reservoir back into place. Make sure that the reservoir is secure to avoid water leakage.
F F i i l l l l i i n n g g t t h h e e r r e e s s e e r r v v o o i i r r w w i i t t h h s s t t e e r r i i l l e e d d i i s s t t i i l l l l e e d d w w a a t t e e r r
The removable reservoir is translucent for you to see the water levels.
C C A A U U T T I I O O N N
• Always use sterile distilled water or distilled water that has been passed through a filter less than or equal to 0.22
microns with this product.
• Always fill the reservoir with room temperature sterile distilled water to reduce the risk of burn.
• Do not overfill the reservoir to avoid the risk of water spillage and fall.
To fill the removable reservoir with sterile distilled water:
1. See Removing and replacing the reservoir (page 26).
2. Fill the reservoir with five liters of sterile distilled water. Do not fill past the top fill line to avoid water overflow (Figure 17).
EN
26