Troubleshooting

Error
Sipper module not
There is no USB
connected. Please
connection between the
check cable.
sipper and the
turbidimeter.
Please check sipper
The pump tubing is not
and tube.
correctly installed.

Replacement parts and accessories

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Description
Examine the USB cable. Make sure that the cable length is
not more than 1 m (3.3 ft). Make sure that the USB cable is
connected to the turbidimeter and the sipper.
Loosen the peristaltic pump cover, then install it again. Refer
to the illustrated steps in
on page 14.
Operate the pump for 15 seconds. Push down the lever and
make sure that the tube is correctly installed around the
rollers.
W A R N I N G
Solution
Replace the pump tubing
English 15