Important safety information
This section contains important safety information
regarding the power cell. For more safety information,
refer to the safety manual and the user guide supplied
with the hearing instruments.
NOTICE
Do not exchange power cells. This may only be
X
performed by Service Center only.
General warnings
WARNING
A power cell (lithium‑ion rechargeable battery)
is permanently built into the hearing instrument.
Incorrect use of the power cell or the hearing
instrument can cause the power cell to burst.
Risk of injury, fire or explosion!
Follow the safety instructions for power cells in
X
this section.
8
Defective power cells
In case you think the power cell is defective comply with
the following steps:
Place the hearing instrument into the charger. If you
X
get a red LED then the power cell is defective.
Perform visual inspection of the hearing instrument.
X
The test is considered failed if it has any of below
criteria:
●
There is physical sign of deformation caused by
bloated lithium power cell inside.
●
There is visible leakage from the lithium power cell.
●
Product is generating heat on its own or hot to the
touch.
If any of the test steps failed, consider the power cell
defective. Due to safety concerns the hearing instrument
cannot be shipped anymore. Please record the serial
number of the hearing instrument, take a photo of the
damages, and consult with Sivantos Customer Service
team before disposal of the product according to local
regulations.