Message
Instrument update file is
missing
Instrument update file is
corrupt
Not enough memory to
update the instrument
USB memory is not
connected.
Read/Write to USB flash drive
Cannot write to USB
memory
Cannot read from USB
memory
Restore backup
No instrument backup is
available.
Not able to restore the
backup
Security
Invalid password
Send data
Connect a receiving
device.
Add sample IDs from list
No valid data found
Not able to read sampling
date.
The instrument cannot
read the Sample ID
Problem/Error: Incorrect
date
Possible cause: The
wrong date format.
The sample ID list full.
Data has not been added.
24 English
Remove the instrument update files from the USB flash drive. Save the
instrument update files again to the USB flash drive.
Connect the USB flash drive to the instrument. Start the instrument
update procedure again.
Contact technical support.
Connect a USB flash drive to the instrument. Make sure that the file
system "FAT32" is installed on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Connect the USB flash drive. Start the instrument update
procedure again.
Connect a USB flash drive to the instrument. Make sure that the file
system "FAT32" is installed on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Look for remaining space on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Connect the USB flash drive to the instrument.
Connect a USB flash drive to the instrument. Make sure that the file
system "FAT32" is installed on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Connect the USB flash drive. Start the instrument update
procedure again.
Enter the correct password. If the password is lost, contact technical
support.
Examine the device connections. Set the Auto Send setting to off.
Refer to
Configure the measurement settings
No sample ID file was found on the USB flash drive.
Make sure that the date and time format is dd.mm.yyyy hh:mm.
Examine the text strings. Refer to the expanded user manual on the
manufacturer's website.
Make sure that the date and time format is dd.mm.yyyy hh:mm.
Remove the sample IDs that are not used. Add a new sample ID.
Solution
on page 12.