Accuracy (cont'd)
Accuracy for K-Type Thermocouple Measurements
Instrument error is ± 1.0% of reading or 1 °C (1.8 °F), whichever is greater. Specification applies at
ambient temperature of 17 °C to 29 °C (63 °F to 84 °F). Thermocouple range or accuracy not included.
Consult thermocouple probe data sheet for usable temperature range and accuracy specification.
Specifications
Display: 4–digit LCD (999.9 or 9999 maximum display)
Response Time: 1 second, 90% response
Temperature Range:
Infrared Sensor: -60 °C to 625 °C (-76 °F to 1157 °F)
K-Type Thermocouple Probe: -64 °C to 1400 °C (-83.2 °F to 2552 °F)
Spectral Response: 8 to 14 µm
Repeatability: ± 0.5% of reading or ± 1 °C (± 2 °F), whichever is greater
Emissivity: 0.10 to 1.0, digitally adjustable
Ambient Operating Temperature Range: 0 °C to 50 °C (32 °F to 122 °F)*
Relative Humidity: 10% to 90% (non-condensing) at < 30 °C (86 °F) ambient temperature
Altitude: 2000 m (6500') maximum
Storage Conditions: -10 °C to 60 °C (14 °F to 140 °F)*.
Remove batteries.
Battery: Two 1.5 volt batteries (AAA, NEDA 24A or IEC LR03)
* For improved accuracy, allow unit 30 minutes to adjust to large change in temperature.
Statement of Conformity
Greenlee Textron Inc. is certified in accordance with ISO 9001 (2000) for our Quality Management
Systems.
The instrument enclosed has been checked and/or calibrated using equipment that is traceable to the
National Institute for Standards and Technology (NIST).
TG-2000