Stryker 242-000-012 Guia Del Usuario página 13

Bandeja de esterilización para 12 instrumentos
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 76
Sterilization
Temperature
Relative Humidity
Vacuum Set Points
EtO Concentration
Time
Wrapping
Temperature
Time
Dry Time
.Warning: Devices with moveable jaws must
be sterilized with the jaws in the open position,
or incomplete sterilization will result.
•. Prepare.the.device(s).and.the.sterilization.tray.
as.indicated.in.the."Sterilization.Tray.Setup".
section.below.
•. Sterilize.the.device(s).inside.the.tray.according.
to.the.parameters.indicated.below.
Ethylene oxide (EtO)
Preconditioning
54°C
(129°F)
70%
1.30.psia
1.hour
Autoclave (Steam)
Prevacuum
Double
132°C
(270°F)
4.minutes
25.minutes
Sterilization
54°C
(129°F)
70%
600.mg/L
(100%.EtO)
4.hours
Prevacuum
Double
134°C
(273°F)
3.minutes
25.minutes
Aeration
55.±.4°C
(131.±.7°F)
12.hours
"Flash"
Prevacuum
132°C
(270°F)
4.minutes
EN-9

Publicidad

Tabla de contenido
loading

Tabla de contenido