Sterilization Tray Setup

Maximum Weight Load
(instruments and tray
combined)
Internal Stacking
External Stacking
Accessories
Instrument Distribution
Autoclave (Steam)
Gravity
132°C (270°F)
Time
10 minutes
40 minutes
* An 18-minute cycle, mandatory in some European countries, is also validated.
Caution: Do not exceed 140°C (284°F) during sterilization, or product
damage may result.
Warning: Trays with non-absorbent liners (such as plastic or silicone-
fingered organizing mats) can cause condensate to pool. To preserve sterility,
remove devices from standing solution.
Warning: Drying time depends on several variables, including: altitude,
humidity, type of wrap, preconditioning, size of chamber, mass of load,
material of load, and placement in chamber. Users must verify that drying
time set in their autoclave yields dry surgical equipment.
Warning: These storage guidelines are not provided as a means to maintain
sterile devices within the tray.
• Between uses, store the tray in a clean, dry environment.
• To prevent damage, minimize contact between the tray and other equipment.
T.A.G. Medical Products Corporation LTD.
Kibbutz Gaaton 25130, Israel
Tel: +972.4.9858400 Fax: +972.4.9858404
6.6 kg (14.6 lbs)
• One internal rack and silicone mat (each provided with this tray).
• Do not stack instruments on top of each other.
Do not stack other trays or devices on or below this tray.
There are no accessories available for use with this tray.
Instruments should be placed in the tray as illustrated in the following pictures.
Prevacuum 1
132°C (270°F)
4 minutes
15 minutes
8
Prevacuum 2
134°C (273°F)
*3 minutes
15 minutes