INSTRUCTIONS FOR USE
Storage Specifications
Storage temperature
Storage Relative humidity range
Operating temperature
Operating Relative humidity range
Electrical Specifications
Not Applicable.
Software Specifications
Not Applicable.
Compatibility Specifications
The C-Prone Head Positioner is
compatible with:
Operating Room Table
Compatibility
Note: Consult the corresponding IFU for the products mentioned in the above table.
Sterilization Instruction:
This device is not intended to be sterilized. Equipment damage may occur.
Cleaning and Disinfection Instruction:
WARNING:
Do not use bleach or products that contain bleach to clean the device. Injury
or equipment damage can occur.
After each use, clean the device with alcohol-based wipes.
Do not put the device into water. Equipment damage can occur.
Use a cloth and a quaternary ammonium disinfecting/cleaning solution to
clean and disinfect the device.
Read and follow the manufacturer's recommendation for low-level disinfection.
Read and follow the cleaning product's instructions. Use caution in areas
where liquid can get into the mechanism.
Wipe the device with a clean, dry cloth.
Make sure that the device is dry before you store it or use it again.
CAUTION: DO NOT IMMERSE PADS IN ANY LIQUID
CAUTION: DO NOT USE BLEACH OR PHENOLICS ON PADS
Document Number: 80028212
Version: A
Description
-29ᵒ C to +60ᵒ C
15% to 85%
This device is intended to be used in a
controlled Operating Room environment.
Description
Not Applicable.
Description
Not Applicable.
Description
Spine System
Jackson Table
Other spinal frame with 17" spaced 1¼"W x 1 ½"H
Comfort Mask Disposable
The C-Prone Head Positioner is compatible with
the following surgical table rail styles: US, UK, EU,
DEN, JP, SWISS.
Page 17
Issue Date: 25 MAR 2020
Ref Blank Template: 80025118 Ver. E