QUICK REFERENCE GUIDE: KI7340C / 7740C SERIES OPTICAL TWO WAY LOSS TEST SET
AUTOTEST OPERATION
To initiate Autotest, connect Power Meter section to Light Source section and
press [Autotest]
To show one wavelength only, scroll [-/+]. Press [Menu] to exit.
Press [Hold] to hold / resume display update.
Autotest remains for 8 seconds after signal loss, to allow quick connection
change.
MEASURING TWO WAY LOSS IN AUTOTEST
Remote reference: on each instrument, connect two way port to power
meter port. Select [Autotest], [Abs/Rel]. Press [Set Ref] for 3 seconds. Meters
display about 0.00 dB.
Local reference: connect two way ports of two instruments. Select [Autotest]
and [Set Ref]. Meters display about 0.00 dB.
Loss testing: after referencing, connect two way ports of both instruments to
test system. Select [Autotest] on one unit. 'Auto' shows while synchronising,
then average loss.
To show dBm/dB results for each direction, press [◄►] successively on
hidden keypad. Press [Menu] to exit.
Two Way Communication: press [Hold] to make remote end beep and flash
[Hold]. Then at Remote end, press [Hold] to resume display update. Local
end beeps and stops [Hold] flashing.
To exit Autotest, remove connection, or press [Menu].
MEASURING ONE WAY LOSS IN AUTOTEST
Reference: connect a two way/source port to a power meter port. Select
[Autotest], [Abs/Rel]. Press [Set Ref] for 3 seconds. Meter displays about
0.00 dB R.
Loss testing: after referencing, connect test system to two way / source
and meter ports. Loss will be automatically displayed.
To exit Autotest, remove test lead for more than 8 seconds or turn off
source.

MEMORY OPERATION:

Data can be stored in all modes, however, memory recall and cancel
functions are not available in Autotest and only work in Light Source,
Power Meter or Return Loss modes.
Clear all memory: press both [MR] and [Exit] simultaneously for 3
seconds. 'clr' will display.
To store in next location, press [M+].
To change next store location, press [◄►] [-/+] [Set] or [Menu].
To recall, press [MR] and [-/+] to scroll memory. Use [Abs / Rel] to see
dB/dBm values. Hold down [Abs/Rel] to see reference level.
Autotest memory recall: use [-/+] to select next wavelength.
Two way test only: press [◄►] repeatedly to scroll details.
To exit memory display, press [Menu].