IRP25.50
2.2
Sensor-/Stylus Change
2.2.1
Sensor Change (Temperature Measuring Unit TP44.10)
Risk of damage to the equipment!
•
Clean and dry the probe well!
•
Do NOT blow off with compressed air!
•
Do not touch the sensor tip.
1.
Solely hold the white base of the socket and pull carefully.
2.
Press sensor completely into the socket and take care of the orientation of the contact plugs.
Fig. 9
Sensor change temperature measuring unit TP44.10
2.2.2
Stylus Change (Measuring Unit MY21.00
Risk of damage to the equipment!
Do not apply any turning force to the measuring unit!
•
During replacement always hold the stylus connection with the Spanner AF4 mm!
1.
Unscrew the stylus from the touch probe using the mounting pin.
2.
Carefully screw the new stylus into the touch probe (Fig. 10).
3.
Align the stylus to spindle center (refer to chapter 2.4).
4.
Calibrate the probe.
Fig. 10
Stylus change measuring unit MY21.00
NOTICE
Plug
NOTICE
Operation
Temperature sensor
Spanner AF4 mm
(1665)
Mounting pin
(0885)
Stylus
M=0,6 Nm (0.44 lbf.ft)
71