Instructions

To Process the Equipment

WARNING: Upon initial receipt and before each
use, process the equipment as indicated in the care
instructions manual supplied with the equipment.

To Charge a Power Module

Fully charge the power module before first and every use.
NOTES:
▪ For charging instructions, see the instructions for use
supplied with the battery charger.
▪ For additional information, including safety information,
see the instructions for use supplied with the power
module.

To Install a Power Module

Sterile Assistant

1. Set the safety switch to the safe mode position.
2. Turn the latch to the unlock position and open the
door.
www.stryker.com
3. Place the sterile transfer shield over the open
compartment.
4. Present the compartment to the Circulating Assistant.

Circulating Assistant

WARNING: DO NOT contaminate the handpiece.
The transfer shield guards the sterile handpiece from
contamination during power module installation.
1. Hold the power module by the handle.
2. Insert the power module through the transfer shield
into the open compartment.