WAVEFORMS (CONTINUED)
MONOPHASIC: Monophasic Rectangular
Pulsed
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
The Monophasic Rectangular Pulsed waveform is an interrupted
unidirectional current with a rectangular pulse shape.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Phase Duration
Phase Interval
IRMS
DC component
MONOPHASIC: Monophasic Triangular
Pulsed
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
The Monophasic Triangular Pulsed waveform is an interrupted
unidirectional current with a triangular pulse shape.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Phase Duration
Phase Interval
IRMS
DC component
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
Galvanic Current is a direct current flowing in one direction only.
The current can be continuous or interrupted.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-60 mA (CC)
Cycle Time
0.1-1,000 ms
Polarity Reversal
5-5,000 ms
0-47mA
Yes
IRMS
DC component
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
Galvanic Current is a direct current flowing in one direction only.
The current can be continuous or interrupted.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-60 mA (CC)
Pulse Duration
0.1-1,000 ms
Phase Interval
5-5,000 ms
Polarity Reversal
0-27mA
Yes
Polarity Reversal Ramp
IRMS
DC component
GALVANIC: Continuous
With Polarity Reversal On, Polarity will change in
the middle of the treatment time.
GALVANIC: Interrupted
With Polarity Reversal On, Polarity will
change in the middle of the treatment time.
E N
Electrodes
1, 2
1-60 min
CC
0-40 mA (CC)
Continuous, or User Defined
On or Off
0-44mA
Yes
Electrodes
1, 2
1-60 min
CC
0-40 mA (CC)
136 μsec
25 usec
On or Off
1 sec
0-41mA
Yes
27