iHealth Air Manual De Funcionamento página 13

Tabla de contenido

Publicidad

Scatter plot of SaO2 versus SpO2 with linear
regression fit
The figure below shows the graphical plot of
SaO2 versus error (SpO2 – SaO2) with upper
95% and lower 95% limits of agreement:
11.Pulse Rate Measuring Range: 30/min-250/min
12.Pulse Rate Accuracy: 30/min ~ 99/min: ±2, 100/min ~ 250/min: ±2%.
13.Data update period: 15s
14.Automatic Shut-off: After 8 seconds of no indication on the sensors
15.Operation Environment: 5℃-40℃; Humidity <80%; Atmospheric pressure:
700hPa-1060hPa
16.Storage Environment: -20℃-55℃; Humidity <95%; Atmospheric pressure:
700hPa-1060hPa
Scatter plot of difference between methods
against the SaO2
11

Publicidad

Tabla de contenido
loading

Productos relacionados para iHealth Air

Este manual también es adecuado para:

Po3m

Tabla de contenido