Mains Adapter; Electromagnetic Compatibility - Chicco TIRALATTE ELETTRICO Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 39
- Relative humidity: 30 to 85%
- Pressure: 700 to 1060 hPa
BATTERY POWERED MODE:
6V DC – 4x1.5V AA size (LR06) batteries

MAINS ADAPTER

Mod. LXCP12-006100BEH, made by: Shenzhen Longxc Power Supply Co., LTD.
INPUT: 100-240V~ 50/60Hz, 0.5A
OUTPUT: 6.0V d.c. 1.0A
Date of manufacture:
The date of manufacture of the product may be found from the batch coding:
(where: YY=year; MM=month; XXXX=progressive number).

ELECTROMAGNETIC COMPATIBILITY

Electro-medical devices require specific precautions in terms of electromagnetic compati¬bility and must be installed and
prepared for use according to the following EMC information. Electromagnetic/Radio Frequency Interference (EMC/RFI):
the behavior of the device during its operating may be subject to interfer¬ence if the Electric Breast Pump is used in an
radiofrequency electromagnetic field with intensity ≤ 10 Volts/m, but the appliance effectiveness will not be permanently
compromised. Keep the Electric Breast Pump at a distance of, at least, 1 m from RF transmitters in order to avoid Radio
Frequency Interferences.
The Electric Breast Pump mod. 06836
below specified. The customer or the user of the Electric Breast Pump mod. 06836
that it is used in such an environment.
Guidance and manufacturer's declaration – Electromagnetic emissions
Phenomenon
Conducted and radiated RF
EMISSIONS
Harmonic distortion
Voltage fluctuations and
flicker
a) The equipment is suitable for use in Home Health Environment and Professional Health Care Environment limited to patient
rooms, breastfeeding and nursery facilities in hospital or clinics. The more restrictive acceptance limits of Group 1 Class B (CI-
SPR 11) have been considered and applied. The equipment is suitable for use in the mentioned environment when directly
connected to the PUBLIC MAINS NETWORK.
b) The test is not applicable in this environment unless the ME EQUIPMENT and ME SYSTEM used there will be connected to
the PUBLIC MAINS NETWORK and the power input is otherwise within the scope of the Basic EMC standard.
Guidance and manufacturer's declaration – Electromagnetic immunity – Enclosure port
Phenomenon
ELECTROSTATIC
DISCHARGE
Radiated RF EM fields
00009199000000 is intended for use in the electromagnetic environment
Professional healthcare facility environment
a
Basic EMC standard or
test method
IEC 61000-4-2
IEC 61000-4-3
24
10YYMMXXXX
00009199000000 should assure
a
)
)
IEC 61000-3-2
b
)
Class A
IEC 61000-3-3
b
)
COMPLY
Livelli prova di immunità
Professional healthcare
facility environment
± 8kV contact
± 2 kV, ±4 kV, ±8 kV, ±15 kV air
a)
HOME HEALTHCARE
ENVIRONMENT
)
a
CISPR 11
Group 1 Class B
HOME HEALTHCARE
ENVIRONMENT
)
a
10 V/m
)
b
80MHz – 2,7 GHz
80% AM at 1kHz