Error message

It is possible for the media to become lodged between the cutters
of the DT-tubes. The unit will automatically change the direction
of rotation up to three times in an attempt to free the cutters. The
unit will switch off automatically if this is not possible.
If the device becomes blocked on three occasions within a certain
period of time the following error message will occur:
 Display:
 An audible signal will be given.
Corrective action: Switch off the device, check
contents, switch on the device.

Accessories

See more accessories on www.ika.com.

Maintenance

The device is maintenance-free. It is only subject to the natural wear
and tear of components and their statistical failure rate.
For cleaning disconnect the mains plug.
Only use cleansing agents which have been recommended by IKA:
Water (containing surfactant) and isopropyl alcohol.
- Wear protective gloves during cleaning the instruments.
- Electrical instruments may not be placed in the cleansing agent for
the purpose of cleaning.
- Do not allow moisture to get into the instrument when cleaning.
- Before using another than the recommended method for clean-
ing or decontamination, the user must ascertain with IKA that
this method does not destroy the instrument.
Spare parts order:
When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
- item and designation of the spare part, see www.ika.com,
spare parts diagram and spare parts list.
Repair
In case of repair the device has to be cleaned and free from
any materials which may constitute a health hazard.
For repair, please request the "Decontamination Certificate"
form IKA or use the download printout of it from IKA website:
www.ika.com.
If you require servicing, return the appliance in its original pack-
aging. Storage packaging is not sufficient. Please also use suitable
transport packaging.

Warranty

In accordance with IKA warranty conditions, the warranty period is
24 months. For claims under the warranty please contact your local
dealer. You may also send the machine direct to our works, enclosing
the delivery invoice and giving reasons for the claim. You will be
liable for freight costs.
The warranty does not cover wearing parts, nor does it apply to faults
resulting from improper use or insufficient care and maintenance
contrary to the instructions in this operating manual.