Failure Remedies - DTF PURENEB AEROSONIC+ Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27
No special knowledge or training is required to use ATOMISOR/PURENEB devices.
Do not touch the device for more than 10 seconds during use.
Do not touch the tubes for more than 1 minute during the device use (their temperature can reach 54 °C maximum
in case of defect).
Carefully keep these instructions for use booklet as it provides necessary information for the good use of the
equipment.
The electric cord cannot be replaced by the maintenance staff.
The patient is an intended operator. He can use all functions of the device safely.
Contact the manufacturer if you need help to set up, use or maintain the medical device and to report any
unexpected operation or event.
Remarks regarding aerosoltherapy
Do not exceed 6 ml medicine in the nebulizing kit.
Do not exceed the mark "full" printed on the nebulizing kit.
If necessary clean the jets of the nebulizer with hot water, without using any metal wire.
Any modification of jets dimensions has a bad effect on particle size and therapeutic aerosols efficiency.
Each nebulizing kit is intended for a single patient.
Aerosols treatment can be done on medical prescription only and within the prescribed doses.
Side effects: for corticoids, it is better to use a mouth piece. Otherwise, if the mask is used, it is recommended to
rinse one's face and/or mouth after the session. Refer to the medicine instructions for use.
Side effects: a risk of allergy of the body parts in contact with the mask or nose/mouth pieces can appear.
Performances supplied by the manufacturer may not apply to drugs in suspension or in very viscous form. In that
case, it is necessary to check with the drug supplier.
Devices not to be used in presence of a mixture of inflammable anaesthetic with air or oxygen or nitrous oxide.
CAUTION, in all cases :
WARNING: Always disconnect the electrical cord from the mains supply before opening the body of the compressor
(electrical shock hazard).
WARNING: Caution, only a professional repair shop is allowed to repair the device.
WARNING: Modification forbidden.
WARNING: Only use the accessories supplied and mentioned in the instructions for use.
WARNING: Do not undertake service or maintenance operations while the device is in use.

8. Failure remedies

Problem
No operating noise can be heard.
No aerosol comes out of the
nebulizing kit.
Check that the electric cord is connected to the mains and to the device.
Check that the switch is in position on.
Check that main power supply is in accordance with the unit (Volt Hz).
Check that electrical current arrives to the socket trying with another device.
Check that the tank contains enough liquid depending on the jet nebulizing kit:
the minimum of medicine is indicated on the instructions for use of each
nebulizing kit.
Check that the tube is well connected to the compressed air inlet of the
nebulizing kit and to the air outlet of the compressor.
Check that the tube is not pinched or blocked.
Check that the jet nozzle is completely inserted.
Jet nozzle must be cleared (clean it with water).
In case of use of medicine glass bulbs, check that glass fragments did not fall
into the nebulizer.