Descargar Imprimir esta página

Maintenance; Clean The Sensor - Hach AV9000 Manual Del Usuario

Ocultar thumbs Ver también para AV9000:

Publicidad

Idiomas disponibles

Idiomas disponibles

3. Subtract the number from step 2 from the pipe diameter
The result is the water depth
4. Use the Adjust Level function of the software to enter the physically-measured water depth.
Figure 7 Measure the water level
1 Water level

Maintenance

Note: Only qualified personnel must conduct the tasks in this section of the manual.

Clean the sensor

Clean the transducer port when:
• Unexpected increases or decreases in flow or level trends occur
• Level data are missing or incorrect but velocity data are valid
• Excessive silt deposits have built up between the transducer and the protective cover
Notes
• Do not touch the sensor transducer as this will cause damage and incorrect sensor operation.
• Use only approved cleaning solutions as listed in
clean the pressure transducer as this will cause damage and incorrect sensor operation. If there
are debris, spray the membrane with water and use a Q-tip to carefully remove the buildup.
• If the gasket is missing or damaged, install a new one. A damaged or missing gasket will cause
inaccurate readings.
• After cleaning the sensor, clean the gasket and protective cover before they are installed.
• After cleaning an oil-filled sensor, replenish the sensor oil. Follow the oil-replenishment procedure
in the instruction sheet included in the Oil Kit (7724800).
• If a sensor must be taken out of service for an extended period, do not store the sensor on a dry
shelf. The manufacturer recommends that the sensor be stored with the sensor head in a bucket
of water to keep the oil debris from crusting in the pressure transducer canal.