The device is maintenance-free. It is only subject to the natural
wear and tear of components and their statistical failure rate.
Cleaning
For cleaning disconnect the mains plug!
Use only cleaning agents which have been approved by IKA to
clean IKA devices.
These are water (with tenside) and isopropanol.
Wear protective gloves during cleaning the devices.
Electrical devices may not be placed in the cleansing agent for the
purpose of cleaning.
Do not allow moisture to get into the device when cleaning.
Before using another than the recommended method for clean-
ing or decontamination, the user must ascertain with IKA that this
method does not destroy the device.
For accessories see www.ika.com.
Any malfunctions during operation will be identified by an error message on the display.
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated.
Proceed as follows in such cases:
- switch off device using the main switch,
- carry out corrective measures,
- restart device.
Error code
Cause
E03
Temperature inside the device
is too high
E04
Target speed is not reached
Motor blocked
E09
Flash memory storage error
If the actions described fail to resolve the fault or another error code is displayed, then take one of the following steps:
- contact the service department,
- send the device for repair, including a short description of the fault.

Maintenance and cleaning

Spare parts order
When ordering spare parts, please give:
- device type.
- serial number, see type plate.
- position number and description of spare part,
see www.ika.com.
- software version.
Repair
Please only send devices in for repair that have been
cleaned and are free of materials which might present
health hazards.
For repair, please request the "Safety Declaration (Decontami-
nation Certificate)" from IKA or use the download printout of it
from IKA website at www.ika.com.
If your appliance requires repair, return it in its original packaging.
Storage packaging is not sufficient when sending the device - also
use appropriate transport packaging.

Accessories

Error Codes

Effect
Room temperature > 40 °C
Load too large
Cable break
Transmission error
Correction
- Switch off the device
- Let the device cool down
- Check the location (see Technical data)
- Reduce the volume in the evaporator
- Turn the unit off and on again