the circuit to normal operation.
Wrong Input Warning Feature
If the display shows
the meter is set up correctly and test the
meter's fuses as described under "Auto Fuse
Detection" in chapter 6. If the rotary switch
is not correctly set to one of the current
measuring positions, the beeper warns you
by making a chirping sound. This warning
is intended to stop you from attempting to
measure other values when the leads are
plugged into a current terminal.
Placing the probes in parallel with a powered
when a lead is plugged into a current terminal
can damage the circuit you are testing and
blow the meter's fuses because the resistance
through the meter's current terminals is so
low that the meter acts like a short circuit.
Auto/ Manual Range Operation
Press the RANGE button momentarily to
select manual-ranging and the meter will
remain in the range it was in when the LCD
annunciator
turns off.
Press the button momentarily again to step
through the ranges. Press and hold the
RANGE button for two seconds or more to
resume auto-ranging.
*Manual-ranging feature is not available in
Hz function.
Beeper
A single beep indicates correct operation.
A double beep indicates a warning or error
condition. Continuous beeping indicates
there is circuit continuity while in continuity
mode.
Auto-Power-Off
The Auto-Power-Off feature has two steps.
The first step has the meter automatically
or
, be sure
go into the power saving mode to extend
battery life after approximately 15 minutes of
no activity. When the meter enters the power
saving mode, the meter beeps warning tones
every minute. To turn on the meter, press
any button or move the rotary switch to any
position.
The second step is to automatically turn the
meter completely off after approximately
15 minutes from the time when the meter
entered the power saving mode. To turn on
the meter after auto-power-off, turn the
rotary switch to any desired position from
OFF position.
You can disable auto-power-off by Setup
Menu. Both the auto-off time and the power
saving mode beep alert time can be adjusted
by Setup Menu.

6. MAINTENANCE

WARNING: TO AVOID ELECTRICAL
SHOCK OR PERSONAL INJURY, REMOVE
THE TEST LEADS AND ANY INPUT SIGNALS
BEFORE REPLACING THE BATTERY OR FUSES.
TO PREVENT DAMAGE OR INJURY, INSTALL
ONLY THE SAME TYPE OF FUSES OR
EQUIVALENTS.
Cleaning and Storage
Periodically wipe the case with a damp cloth
and mild detergent; do not use abrasives or
solvents.
Clean the input terminals as follows:
1. Turn the meter off and remove all test
leads.
2. Shake out any dirt that may be in the
terminals.
3. Soak a new swab with alcohol and work the
swab around in each terminal.
If the meter will not to be used for periods
longer than 60 days, remove the battery and
store it separately.
E N G L I S H