Electromagnetic Compatibility

Like other electrical medical equipment, the 32" 4K Surgical Display requires special precautions
to ensure electromagnetic compatibility with other electrical medical devices. To ensure
electromagnetic compatibility (EMC), the display must be installed and operated according to the
EMC information provided in this manual. The display has been designed and tested to comply
with IEC 60601-1-2 requirements for EMC with other devices.
Warning
Caution
Guidance and Manufacturer's Declaration: Electromagnetic Emissions
The 32" 4K Surgical Display is intended for use in the electromagnetic environment specified below. The customer or the user
of the display should ensure it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage Fluctuations/ flicker
emissions IEC 61000-3-3
EN-24
When this device is connected with other electrical equipment,
leakage currents may be additive. To minimize total leakage current
per patient, ensure that all systems are installed according to the
requirements of IEC 60601-1-1.
Portable and mobile RF communications equipment may affect the
normal function of the display.
Do not use cables or accessories other than those provided with the
display, as this may result in increased electromagnetic emissions or
decreased immunity to such emissions.
If the display is used adjacent to or stacked with other equipment,
observe and verify normal operation of the display in the configuration
in which it will be used prior to using it in a surgical procedure. Consult
the tables below for guidance in placing the display.
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic Environment - guidance
The display use RF energy only for their internal function;
therefore, their RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The display is suitable for use in all establishments other than
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following warning is
heeded:
Warning: This system is intended for use by health care
professionals only. This system may cause radio interference
or may disrupt the operation of nearby equipment. It may be
necessary to take mitigation measures, such as reorienting or
relocating the system or shielding the location.