Cleaning And Disinfection - cleanAIR Shigematsu GX02 Instrucciones De Uso

Ocultar thumbs Ver también para Shigematsu GX02:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 53
Instructions for use
Use a hexagonal wrench to remove the two
1.
screws on the front side and then remove
the front cover.
Use the screws to fasten a new front
2.
cover.
Headband
(see Figures Annex 8.1 - 8.4)
Pull the two pawls of the headband from
1.
the upper section of the buckle.
To release the buckle turn it on its pin by
2.
90° clockwise and counter-clockwise.
Slide the buckle off to remove the head
3.
piece.
4.
Replace the pin.
5.
Install a new headband.

8. Cleaning and disinfection

WARNING
Perform cleaning only when the filter is not installed.
Wash the mask with lukewarm water and mild
detergent. (in case of heavy dirt) use a sponge
or a soft brush for the cleaning.
Thoroughly wash the mask under running wa-
ter.
Perform disinfection using disinfecting agents
diluted in water (max. 30 °C)
Observe the instructions stated in the disin-
fec-ting agent instructions for use.
Thoroughly wash the mask under running wa-
ter.
For service centres
Before cleaning and disinfecting:
Remove the inlet valve, the inlet valve membra-
ne and the outlet valve
Remove the inner mask (including the inner
mask valve and the inner mask valve membra-
ne)
Removed parts must be disinfected separately.
Dry the mask and its components (temperatu-
re max. 50 °C)
Assemble the mask in the reverse order. Af-
ter assembly an air-tightness test should be
performed.
NOTICE
Do not use dishwashers or dryers.
Do not use acetone or other cleaning solvents.
9. Air tightness
WARNING
Air-tightness should be tested in an authorised
service centre.
Air-tightness test must be performed by a testing
method specified in ČSN EN 136.
DANGER
Leaking masks must not be used.
10. Storage
Keep the mask at a suitable clean, dry and shady
place in a temperature range from 0 °C to +50 °C
NOTICE
Do not pile or stack the masks.
Respirators must be disposed of in acordance
with local regulations for the disposal of waste
of this kind.
11. Shelf life
No specific shelf life has been determined for the
masks and their parts.
For filter shelf life see the filter user manual.
12. Disposal
Respirators must be disposed of in accordance
with local regulations for the disposal of waste of
this kind.
GX02 Full-Face Mask
ENG
7

Publicidad

Tabla de contenido
loading

Tabla de contenido