Replacement fuse

To check the instrument FUSE", remove it with a screw driver. If the fuse is rup-
tured replace it by another with the following specifications:
Fuse Schurter, model SPT 5x20 (Time-lag) 2A/250V. High breaking capacity H.

Cleaning

Cleaning of this instrument should be carried out using a soft cleaning anti-static
liquid, after verifying that it doesn't affect the plastic parts used in the case and in
the Control Panel of this equipment.

Technical specifications

Test voltages
Insulation test up to
Test
Voltage
A
1 kV
0 - 20
5 kV
0 - 100
10 kV
0 - 200
15 kV
0 - 300
Short-circuit current
Test voltages accuracy
Insulation tester accuracy : Class 2 (±2% of full scale deflection)
Analogue indicator
Safety class
Environmental protection : IP-54 (with closed lid)
: 1 kV – 5 kV – 10 kV – 15 kV
: 3.000.000 MΩ
MEASURING INTERVALS (MΩ)
B
10 - 300
200 - 20.000
50 - 1.500
1.000 - 100.000
100 - 3.000
2.000 - 200.000
150 - 4.500
3.000 - 300.000
: 1 mA
: ±2% of nominal test voltages @ R ≥ 10 GΩ
: Up to 98 mm scale length, taut band, with mirror (thus
avoiding parallax errors)
: Meets the requirements of IEC 61010-1/1990, IEC
61010-1/1992 amendment 2
28
C
C x 10
2.000 - 200.000
10.000 - 1.000.000
20.000 - 2.000.000
30.000 - 3.000.000
Scale
Output
Multi-
Resis-
plier
tance
x1
1 MΩ
x5
5 MΩ
x10
10 MΩ
15 MΩ
x15