Problem
Possible Solution
E1 error code
E2 error code
"Please
Reboot"
E4 error code
Touchscreen
freezes on
"1488" splash
screen when
powering on
the console
"Restart
Camera
Console"
message
(Color bar
background)
"System Error"
message
(Light blue
background)
All manuals and user guides at all-guides.com
• Console shut down due to main board error.
• Turn off the console, wait 3 seconds, and turn it
back on.
• Console shut down due to digital board error.
• Turn off the console, wait 3 seconds, and turn it
back on.
• Turn off the console, wait 3 seconds, and turn it
back on.
• Maximum operating temperature of console
exceeded.
• Ensure vent holes on console are not obstructed.
• Error occurred during console startup.
• Turn off the console, wait 3 seconds, and turn it
back on.
• Camera head temporarily shut down due to
overcurrent. Turn off the console, wait 3 seconds,
and turn it back on.
• After sterilization, ensure the camera head has
cooled down before connecting it to the console.
• No video detected.
• After sterilization, ensure the camera head has
cooled down before connecting it to the console.
• Return the system for repair.