Cleaning And Disinfection; Accessories, Detachable Parts And Other Supplies; Lifetime; Technical And Performance Data - icare HOME TA022-088-3.1 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 32
www.icaretonometer.com
Instructions for replacing the probe base:
• Turn off the tonometer
• Unscrew the probe base collar and put it in a safe place
• Remove the probe base by tilting the tonometer downwards and use your
fingers to pull the probe base out of the tonometer
• Insert a new probe base into the tonometer
• Screw the collar in, to lock the probe base

15. CLEANING AND DISINFECTION

The forehead and cheek supports must be cleaned for each new patient.
The health care professionals should use a wipe dampened with a 70-100%
isopropyl alcohol solution. The patients should use a cloth or paper towel
dampened with water. Do not immerse the tonometer in water or other liquid.
The tonometer must not be immersed or cleaned using too much water.

16. ACCESSORIES, DETACHABLE PARTS AND OTHER SUPPLIES

Part number Product Description Weight Dimensions
540
Probe base
TA022-118
TA022 narrow collar
560
Wrist strap
TA022-044
Carrying case
7179
Battery cover
571
Battery 3 V, CR123A
TA022-037
Support position tags 40g
575
USB cable
113
Probe Icare TP022,
50 pcs/box

17. LIFETIME

The expected service life of the device is 5 years. The maintenance described
in the chapters 12-15 is required during the the expected service life.
The shelf life of the probes in their intact original packaging is 3 years.
Please check the shelf life (expiry date) from the probe label. We recommend
that you inspect the device for mechanical and functional damage and the
safety labels for legibility annually/every 12 months. The life time of the
batteries is about 5000 measurements.
Applicable in Germany only: Messtechnische Kontrolle nach MPG
(Medizinproduktegesetz) alle 24 Monate.

18. TECHNICAL AND PERFORMANCE DATA

Type: TA022
Dimensions: approximately 11 cm x 8 cm x 3 cm
Weight: approximately 150g
Power supply: 2 x CR123 non-rechargeable batteries (make sure you use
batteries with built-in PTC protection, for example Energizer Lithium Photo
123 3V CR123A)
Measurement range: 5-50 mmHg
Accuracy (95% tolerance interval relatively to manometry): ±1,2 mmHg
(≤20 mmHg) and ±2.2 mmHg (>20 mmHg)
Repeatability (coefficient of variation): <8%
The serial number is located on the inside of the battery compartment cover.
The lot number of the probes is on the side of the probe box and the blister
packing.
There are no electrical connections from the tonometer to the patient. All the
parts of the tonometer are applied parts and the tonometer has BF-type
electric shock protection.
4g
38 mm × 7 mm
2g
20 mm × 13 mm
3g
270 mm × 10 mm × 10 mm
210g
270 mm × 135 mm × 60 mm
3g
26 mm × 23 mm × 7 mm
17g
35 mm × 17 mm
70 mm × 41 mm × 13 mm
23g
1 m
55,14g
195 mm × 82 mm × 35 mm
9

Publicidad

Tabla de contenido
loading

Productos relacionados para icare HOME TA022-088-3.1

Tabla de contenido