IV - INSTALLATION
4. 1 UNPACKING THE DEVICE
On receipt of the device, check for any damage
caused in transit.
If necessary, contact your supplier.
4. 2 RECOMMENDATIONS
Have connections to the electrical and water
supply carried out by an approved dental
installation engineer.
The S
P5 N
should be connected to
®
UPRASSON
EWTRON
the electrical supply in accordance with current
standards. Use of a mains plug with connection to
the earth is compulsory.
The water supply system should satisfy the
quality criteria required for the practice of
dental treatment. In addition, it should be
equipped with a water supply cut-off system.
4. 3 INSTALLATION
- Set up the control unit in the desired position.
- Check that the control potentiometer is set to
position OI (off).
- Connect the S
P5 N
UPRASSON
EWTRON
with its filter to the water supply.
- Connect the mains plug to an electrical socket.
10
- Placethe footswitch so as to make it easy to
operate with the foot.
- Check there is no trace of damp in the
handpiece connection, otherwise dry it (wipe it
and blow on it with the multifunction syringe).
- Connect the handpiece simply by plugging it
into the connector cord, aligning the electrical
contacts.
- Place the handpiece on the stand.
IMPORTANT: do not twist the handpiece
connector on its cord.
4. 4 COMMISSIONING
Before using the S
first time, we recommend following the
instructions below:
- Before the first use, it is vital to clean,
decontaminate and sterilize the accessories (cf
section 6).
- After installation, operate the spray for a few
minutes in order to rinse the device's internal
water circuits: with the device operating at the
lowest power (1) and with pressure being
applied to the footswitch.
- Disconnect the handpiece.
- Carry out maintenance and/or sterilization of
all equipment according to the procedures
defined in chapter 6.
's water tube
®
P5 N
for the
®
UPRASSON
EWTRON