V - USE
The S
P5 N
is operated in the
®
UPRASSON
EWTRON
following manner:
- Check that the device is plugged in and
insulated correctly.
- Check there is no trace of damp in the
handpiece connection, otherwise dry it (wipe it
and blow on it with the multifunction syringe).
- Plug in the handpiece.
- Gently operate the spray: operation should be
at the lowest power (1) and with pressure
being applied to the footswitch.
- Install the required tip.
- Tighten it carefully with the wrench to ensure
optimum operation of the ultrasound (an
"Easytorque F01400" torque wrench is available
as an option).
- Adjust to the desired power (according to the
colour and no.) using the control button. Refer
to the Tipbook to find out the power settings
for SATELEC
tips.
®
important:
- Do not disconnect the cord handpiece when
power to the device is switched on and the
footswitch depressed.
- The S
P5 N
should not be used
®
UPRASSON
EWTRON
with accessories other than those supplied by
SATELEC
.
®
- Correct vibration of a tip depends on correct
tightening which should not be forced beyond
its stop point. In addition, to avoid self-
locking, the tip should be removed after each
use.
- Over-tightening the tip may lead to either it or
the handpiece breaking.
- At the end of the working day:
• the water supply should be switched off;
• the device should be switched off (OI):
• the mains plug should be unplugged.
11