Connection side view
4
5
3
2
1
CONTROL connec-
tion
Connection side view
Connect the gages with a pre-assembled measuring cable (see sales documentation)
or with a self-made, shielded cable (EMC compatibility) to one of the six connections
SENSOR 1 ... SENSOR 6 on the rear panel.
Pay attention to the list of usable measuring tubes!
Dangerous contact voltage
Voltages above 30 V AC or 60 V DC are deemed to be dangerous in the event of contact
in accordance with EN 61010.
Only apply a grounded protective extra-low voltage (PELV).
Connector pin arrangement
The 6-pin C91B device sockets are assigned as follows:
Pin
Signal
1
Identification
6
2
Power supply grounding GND
3
Signal input (measuring signal 0 ... +10 V DC)
4
Signal grounding (measuring signal -)
5
Shield
6
Gage power supply +24 V DC
This terminal can be used to:
● read out the measuring signal,
● analyze the fault monitoring condition in a potential-free fashion,
● switch on/off the gages.
Dangerous contact voltage
Voltages above 30 V AC or 60 V DC are deemed to be dangerous in the event of contact
in accordance with EN 61010.
Only apply a grounded protective extra-low voltage (PELV).
Connect peripheral components to the CONTROL connector on the rear panel using
a self-made shielded connection cable (EMC compatibility).
Connector pin arrangement
The 15-pin high-density D-Sub device socket is assigned as follows::
Pin
Signal
1
Analog output sensor 1 (0 ... 10 V DC)
2
Analog output sensor 2 (0 ... 10 V DC)
3
Analog output sensor 3 (0 ... 10 V DC)
4
Analog output sensor 4 (0 ... 10 V DC)
5
Analog output sensor 5 (0 ... 10 V DC)
6
Analog output sensor 6 (0 ... 10 V DC)
7
GND
8
GND
9
GND
10
External control sensor 1
11
External control sensor 2
12
External control sensor 3
13
External control sensor 4
14
External control sensor 5
15
External control sensor 6
DANGER
DANGER
Installation
7