798 MPT Titrino
Wrong sample. For SET with preset titration direction the first
measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
No EP set for SET.
Exit: The error message disappears on next startup.
Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
Silo full.
Exit: Send new command.
No method. A method, which is required from the silo memory or in TIP, does
not exist.
Exit: The error message disappears on next startup.
Check temp.sensor in MEAS T or with activated temperature monitoring.
Exit: Correct error or &m $S.
Same buffer in CAL. Measured value of the second buffer differs less than 6
mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
XXX Bytes are missing so that the method, the silo line could not be stored or
not enough RAM for running TIP.
Exit: Send new command.
No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
No sequence defined in TIP.
Exit. The error message disappears on next start.
A second TIP has been called up in TIP.
Exit: The error message disappears on next start.
No new temporary variable.
Exit: The error message disappears on next start.
Measurement range of the secondary measured value (temperature)
exceeded. The primary measured value (pH, U, Ipol, Upol) can also be
unstable.
Exit: Rectify error or &m $S.
Save lines is "OFF" although a submethod of TIP includes an assignment to
C24 or C25.
Exit: The error message disappears on next start. Attention: The data of this
sample will not be stored.
In TIP, a QuickMeas was started, without defining a measuring quantity.
Exit: The error message disappears on next start or &Mode.QuickMeas $S.
Accessing to the memory card, the card was not (properly) inserted.
Exit. The error message disappears on next start.
The date of changing the battery of the card is expired.
Exit. The error message disappears on next start.
Memory card write-protected.
Exit: Send new command.
Memory card not formatted.
Exit: Send new command.
Memory card not accessible.
Exit: Send new command.
A directory with the same name exists already on the memory card.
Exit: Send new command.
3.1 General rules
75