Specifications - WIKA 232.50.100 Manual De Instrucciones

Manómetro modelo 2, ng 100 y ng 160 según atex
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3. Specifications

3. Specifications
Pressure limitation
EN
Models 232.50, 233.50, 232.30, 233.30, 262.50, 263.50, 262.30, 263.30,
232.53, 232.54, 233.53, 233.54, PG23LT, PG23CP , PG28:
Steady:
Fluctuating:
Short time:
Models 232.36 and 233.36:
Steady:
Fluctuating:
Short time:
Temperature effect
When the temperature of the measuring system deviates from the refer-
ence temperature (+20 °C): max. ±0.4 %/10 K of full scale value
Case ingress protection
Model 2xx, PG23CP, PG28: IP65, IP66
Model 2xx.3x and back mount: IP54
Model PG23LT for scale range > 0 ... 16 bar: IP66 / IP67
Model PG23LT for scale range ≤ 0 ... 16 bar: IP65
For further specifications see WIKA data sheets PM 02.02, PM 02.04,
PM 02.15, PM 02.22, PM 02.24 and/or PM 02.32 and the order
documentation.
1) For general use, no ATEX requirement
20
Full scale value
0.9 x full scale value
1.3 x full scale value
End value of measuring range
0.9 x end value of measuring range
Overload range
(per IEC/EN 60529)
1)
WIKA operating instructions pressure gauge, model 2 per ATEX

Publicidad

Tabla de contenido
loading

Tabla de contenido