Guidance and manufacturer's declaration
The thermometer are intended for use in the electromagnetic
environment specified below. The customers or the users of these
thermometer should assure that it is used in such environment.
Immunity
IEC 60601
test
Test level
Conducted
RF
Not
IEC 61000
applicable
-4-6
Radiated
3 V/m
RF IEC
80 MHz to
61000-4-3
2.5 GHz
– electromagnetic immunity
Compliance
level
3 V/m
20
Electromagnetic
environment – guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the thermometer including
cables, than the recommended
separation distance calculated
from the equation appropriate to
the frequency of the
transmitter.
Recommend separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to he
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters as determined by an
electromagnetic site survey,