Applied Press.
(mm)
2.0 ATM
3.51
3.0 ATM
3.59
4.0 ATM
3.67
5.0 ATM
3.74
6.0 ATM
3.80
7.0 ATM
3.87
8.0 ATM
3.94
9.0 ATM
4.00
10.0 ATM
4.07
11.0 ATM
4.12
12.0 ATM
4.18
13.0 ATM
4.23
The figures in bold face represent the balloon diameter @ Rated Burst Pressure.
FOR ALL NuMED CATHETERS AN INFLATION DEVICE WITH PRESSURE GAUGE SHOULD BE USED.
COEfficient™ Balloon Sizing Chart
4.0
5.0
6.0
(mm)
(mm)
4.24
5.24
4.38
5.41
4.52
5.47
4.64
5.60
4.76
5.72
4.88
5.83
4.99
5.94
5.07
6.03
5.15
6.14
5.21
6.19
5.27
6.28
7.0
8.0
9.0
(mm)
(mm)
(mm)
6.08
6.99
7.82
6.24
7.13
8.05
6.37
7.35
8.28
6.54
7.50
8.53
6.68
7.70
8.76
6.82
7.89
8.97
6.96
8.07
9.15
7.08
8.20
9.31
7.18
8.33
9.45
7.27
4
10.0
11.0
12.0
(mm)
(mm)
(mm)
8.53
9.69
10.24
8.84
9.87
10.53
9.11
10.06
10.91
9.37
10.38
11.32
9.63
10.72
11.72
9.82
11.02
12.03
9.94
11.23
10.02
10.12