Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

8
To do this, first measure blood pressure on both arms and
subsequently always on the arm which produced the higher
reading.
Place cuff in such a manner that the lower cuff edge finishes
ca. 2 – 3 cm above the elbow and the centre of the rubber
bag rests on top of the artery.
Taking the Blood Pressure:
Inflate the cuff to above the systolic pressure. The air release
speed can be adjusted by rotating the valve screw.
The World Health Organisation (WHO) recommends a
deflation speed of 2 – 3 mmHg per second. To avoid a rea-
ding error, read the pressure values perpendicular to the
scale.
After the measurement is completed, fully open the ai
release valve to allow for the fast deflation of the cuff.

Technical Data:

Measurement range: 0 – 300 mmHg
Accuracy of Pressure indication: ± 3 mmHg
or 2% of the reading (whichever is greater)
Storing conditions:
–30°C – 70°C, 15 – 85% rel. humidity
Operating conditions:
10°C – 40°C, 15 – 85% rel. humidity
Cleaning of Unit and Cuff:
– To clean the unit only use a dry, soft cloth.
– To clean the cuff – remove the rubber bag. The exterior
of the cuff can then be washed by hand at a maximum
temperature of 30 degrees Celsius.
Disinfection:
For disinfectant wipes (at least 5 minutes exposure time) of
the device and the cuff, we recommend the disinfectant mi-
krozid sensitiv liquid (Schülke & Mayr).
To disinfect the cuff, we recommend spray disinfection.

Publicidad

Tabla de contenido
loading

Tabla de contenido