Guidance and manufacturer's declaration- electromagnetic immunity-
Guidance and manufacturer's declaration- electromagnetic immunity-
For EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
Guidance and manufacturer's declaration- electromagnetic immunity
The model MD300K2 Pulse Oximeter is intended for use in the electromagnetic environment specified
below. The customer or the user of the model MD300K2 Pulse Oximeter should assure that it is used in
such an environment.
Immunity test
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
a ected by absorption and reflection from structures, objects and people.
For EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
IEC 60601
test level
IEC 60601 test level
3Vrms 150kHz to
80kHz
3V/m 80MHz to
2.5GHz
Compliance
level
Compliance level
3V
3V/m
- 29 -
Electromagnetic
environment-guidance
Electromagnetic environment-
guidane
c o m m u ni c ations e q u ip m ent
sho u ld be u sed no c loser to
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
c ables , than the re c o m m ended
model MD300K2 Pulse Oximeter,
separation distan c e c al c u lated
fro m
the e q u ation appli c able to
including cables, than the re-
commended separation distance
calculated from the equation
Recommended separation
distance
applicable to the frequency of
the transmitter. Recommended
3.5
separation distance
d=
P
V
1
d=1.2 √P
150KHz to 80 MHz
3.5
d=
P
E
1
d=1.2 √P
80MHz to 800MHz
7
d=
P
E
1
d=2.3 √P
800MHz to 2.5GHz
po w er ratin g of the trans m itter
Where P is the maximum output
trans m itter m an u fa c t u rer and d
power rating of the transmitter in
is the re c o m m ended separation
Watts (W) according to the trans-
mitter manufacture and d is the
recommended separation distan-
ce in meters (m). Field strengths
trans m itters , as deter m ined
by an ele c tro m a g neti c site
from fixed RF transmitters, as
a
s u r v ey ,
sho u ld be less than
determined by an electromagnetic
the c o m plian c e le v el in ea c h
site survey,a should be less than
b
the compliance level in each fre-
quency range.b Interference may
I nterferen c e m ay o c c u r in the
v i c inity of e q u ip m ent m ar k ed
occur in the vicinity of equipment
w ith the follo w in g sy m bol :
marked with the following symbol:
MD300K2
b