Symbols
Appears during measurement when
the cuff is attached loosely
Detected rate of IHB in memory
User
Previous measurements stored in memory.
Average data
The battery power indicator during
measurement.
Full Battery
The battery is low when it blinks.
Low Battery
Unstable blood pressure due to
movement during measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
The pulse was not detected correctly.
PUL. DISPLAY ERROR
Blood pressure monitor internal error
SYS.
Systolic blood pressure in mmHg
DIA.
Diastolic blood pressure in mmHg
PUL./min.
Pulse per minute
AM
Data taken between 4:00 and 9:59
PM
Data taken between 18:00 and 1:59
Symbols
Function / Meaning
English 6
Recommended Action
The reading may yield an
incorrect value. Apply the
cuff correctly and take
another measurement.
Replace all batteries with new
ones when the mark blinks
Take another measurement.
Remain very still during
measurement.
Apply the cuff correctly and
take another measurement.
Remove the batteries and
press the
button,
START
and then install the batteries
again. If the error still appears
contact the manufacturer.