WARNING To avoid potential serious injury to the user and the patient
and/or damage to this device, please note the following
warnings:
1.
To avoid potential serious injury to the user and patient and/or damage
to this device, read the Visum Surgical Light and StrykeCAM In-Light
Camera operating manual thoroughly and be familiar with its contents
prior to using this equipment.
2.
The Visum Surgical Light , StrykeCAM In-Light Camera and the Visum/
StrykeCAM EDAM (hereafter referred to as "EDAM") is designed to be
used by a qualified physician/technician familiar with its use. Misusing
the Visum Lights/StrykeCAM Camera may cause injury to the patient
and/or damage system components.
3.
Carefully unpack the kit and check if any damage occurred during
shipment. If damage is detected, DO NOT use the equipment.
WARNING To avoid potential serious injury to the user and the patient
and/or damage to this device, please note the following
electrical warnings:
1.
To prevent electrical shock to the patient and/or operating room
personnel, do not immerse the Visum Power Box or the EDAM or allow
fluid to enter the chassis or connector.
2.
Before and after each use, inspect the cord insulation for cracks, nicks,
cuts, dents, or depressions. Any irregularity will decrease the insulation
effectiveness and may result in shocks to the patient and/or operating
room personnel.
3.
To prevent injury or electrical shock to the patient and/or operating
room personnel, always power Off the Visum Power Box when
connecting or disconnecting accessories. Caution: To avoid potential
damage to this device, always power Off the Visum Power Box before
connecting or disconnecting the Sidne™ interface kit.
3