14. ELECTROMAGNETIC COMPATIBILITY
(EMC) TABLES
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The CEFAR BASIC Electrotherapy System is intended for use in the electromagnetic environment deified below. The customer or
the user of the CEFAR BASIC Electrotherapy System should assure that it is used in such an environment
Emissions Tests
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not Applicable - Battery powered
IEC 61000-3-2
Voltage fluctuations
Not Applicable - Battery powered
IEC 61000-3-3
Compliance
The CEFAR BASIC Electrotherapy System uses RF energy only
for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.
CEFAR BASIC Electrotherapy System is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes
Electromagnetic Environment - Guidance