Descargar Imprimir esta página

Set To The Factory Default Calibration - Hach DR1300 FL Manual Del Usuario

Ocultar thumbs Ver también para DR1300 FL:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27
5. Select CAL-SLP.
6. Measure the Standard 1 vial as follows:
a. Enter the concentration for the Standard 1 vial (ppb) from the
Certificate of Analysis for the parameter selected in step 4.
b. Clean the Standard 1 vial with a no-lint cloth.
c. Install the Standard 1 vial in the vial adapter.
d. Put the vial cover on the sample vial.
e. Push Enter to measure the Standard 1 vial.
7. Measure the Standard 2 vial as follows:
a. Enter the concentration for the Standard 2 vial (ppb) from the
Certificate of Analysis for the parameter selected in step 4.
b. Clean the Standard 2 vial with a no-lint cloth.
c. Install the Standard 2 vial in the vial adapter.
d. Put the vial cover on the sample vial.
e. Push Enter to measure the Standard 2 vial.
8. Push Enter to save the results.

8.4 Set to the factory default calibration

To remove a user-entered slope calibration from the instrument and
use the factory slope calibration, do the steps that follow:
1. Select CALIBRATION, then select the parameter for the
calibration.
2. Select DEFAULT, then push Enter to confirm.
Section 9 Data management
9.1 Show the measurements on the instrument
1. Select SYSTEM > Log.
2. Selete the month the measurements were done, then push Enter.
3. Push the DOWN and UP arrows to scroll through the
measurements.
English 19

Publicidad

loading

Este manual también es adecuado para:

Lpv449.97.01002Lpv449.98.01002Lpv449.99.00002