The GIMA Peep Valve (Accessory Series)
Terminal Positive Pressure Adjustment Valve
(for one patient use)
1. 2-10 cmH
O Terminal Positive Pressure Adjustment Valve for one patient use
2
2. 5-20 cmH
O Terminal Positive Pressure Adjustment Valve for one patient use
2
3. Terminal Positive Pressure Adjustment Valve Adaptor
Terminal Positive Pressure Adjustment Valve
(article name, number and specification)
2-10 cmH
O Adjustable Terminal Positive Pressure Adjustment Valve (Orange Silicone)
2
5-20 cmH
O Adjustable Terminal Positive Pressure Adjustment Valve (Blue Silicone)
2
(PM-0050) Terminal Positive Pressure Valve Adaptor
Specification
Adjustable range:
2-10 cmH
(+/-2 cmH
O) Flowing Capacity Adjustment 3 Lpm
2
Adaptor: 30 mm 22/15 mm outer dimension 22mm and 30mm
Materials: PC, Silicone, Stainless Steel
The Durable Peep Valve (User's manual)
1.Attach the Diverter to the patient's outlet as showed in the drawing.
2. Turn the Diverter toward the direction away from the patient or the Emergency Personnel's position.
3. Compress the Resuscitator a few times to make sure all functions are normal after assembly.
4. Choose the proper Durable Peep Valve within the specification range. (2-10 cmH
5. Turn the knob of the Durable Peep Valve to the needed manometer indicated on the valve base.
6. As showed in the drawing, attach the Durable Peep Valve to the Diverter, connect the Resuscitator
patient's outlet to the manometer and the breathing bag, press on the Resuscitator for the air exchange
motion of the breathing bag, and adjust for the proper Durable Peep Valve pressure needed on the
Durable Peep Valve.
7. Regularly clean and sterilize before and after use.
ENGLISH
O and 5-20 cmH
2
O
2
26
O or 5-20 cmH
O)
2
2