We do not accept liability for injury to persons or damage to property if the
safety advice has not been followed.
ProMinent
®
TAKE CARE
• Use the measuring only cell for the purpose of
determining free chlorine or cyanuric acid chlorine in
swimming pool water or water of similar quality.
• To guarantee the flow parameters, the sensor should be
used only in the following in-line probe housing models:
DLG III A (914955), DLG III B (914956) or DGM (25 mm
module).
We cannot accept liability if any other type of inline probe
housing is used.
• The power supply to the measuring device (and the
sensor) must not be interrupted. If the power supply is
interrupted for a long time (> 2 h), the device should be
recommissioned (see chapter 7.1).
• Connect the CGE 2-4P-10 ppm measuring cell only to
ProMinent controllers equipped with a corresponding
connection, e.g. DULCOMETER D_4a for chlorine.
CAUTION
• When installing this equipment abroad, the user must
comply with local regulations for maintenance and
calibration periods.
• Wear safety goggles and protective clothing when
handling water or solutions that contain chlorine.
• Ensure the measuring cell and its peripherals are only
operated by specifically trained and authorized operating
personnel!
• Do not operate the measuring cell outside the specified
measuring range!
• Do not operate the measuring cell in water containing
surfactants.
• Avoid skin contact and swallowing the electrolyte. The
electrolyte is slightly toxic and can be harmful to health.
• Avoid getting air bubbles in the water sample as this
causes incorrect dosage.
Air bubbles adhered to the diaphragm of the measuring
cell can cause a measured value that is too low and thus
lead to incorrect metering.