Joerns Healthcare Oxford Elevate Manual De Instrucciones página 23

Ocultar thumbs Ver también para Oxford Elevate:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

English
*THESE CHECKS SHOULD INCLUDE:
1. BOOM: Check the attachment of the boom to the mast. Make sure there is
only minimal side movement of the boom and the boom is free to rotate on
the boom bearing.
2. SLING ATTACHMENT POINTS: Examine the sling hooks and attach-
ment clips on the spreader bar and boom for excessive wear.
3. MAST: Check the operation of the mast-locking device. Make sure the
mast fully engages into the socket.
4. ACTUATOR: The actuator should require no maintenance other than
checking for correct operation and listening for unusual noise.
5. KNEE PAD: Check knee pad adjuster is fastened securely.
6. SMART MONITOR: Check the function of the emergency stop button.
Inspect the hand control socket for correct fitting. Check functioning of the
hand control. Check the redundant controls and confirm they operate as
intended. The MINIMUM load required to manually lower the hoist using
the emergency lower function is 30 kg.
7. BATTERIES: The batteries are housed in the power pack and should
not require maintenance other than the regular charging as detailed in the
charging instructions. Check that the connections remain clean.
8. BATTERIES (Charging): Confirm the lift is not giving a low battery alarm
when the hand control is operated. If the alarm sounds, do not use and
place the battery on charge immediately.
9. LEG ADJUSTMENT: Check the legs operate in both full extensions
(inward/outward).
10. CASTORS: Check all castors for firm attachment to the legs. Check for
free rotation of the castor and the wheels.
11. CLEANING: Using a damp cloth, wipe clean with ordinary soap and water
and/or any hard surface disinfectant. Harsh chemical cleaners or abrasives
should be avoided as these may damage the surface finish of the lift. Avoid
wetting any of the electrical parts.
12. LOAD TEST: The load test should be carried out in accordance with
the manufacturers test procedures and the directive detailed in EN ISO
10535:2006 - Annex B - Periodic Inspection B1 - see excerpt below. It is
strongly recommended that an authorised service dealer carry out the test.
13. BASE AND WHEELS: Ensure base is even and level (all four wheels are
on the floor).
14. SLINGS: Check for wear and fraying.
15. LUBRICATION: Lubricate pivot joints with a food grade spray lubricant,
including mast and boom connections, pedal assembly, spreader bar joint
(only if required).
16. HAND SET: Ensure plugged fully into controller.
Oxford
®
Elevate
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
ü
23

Publicidad

Tabla de contenido
loading

Tabla de contenido