9.

OPERATOR WARNING

Event
Low battery
Critically low battery
Detecting a child placed on the pad
(evaluated in standby mode)
Detecting a child removed from the
pad (evaluated in active mode)
Low temperature indication
(evaluated in active mode)
High temperature
indication
(evaluated in active mode)
Indication that the sensor pad is
disconnected in standby mode
(evaluated in standby mode)
Mechanical button jam detected in
standby mode (evaluated in standby
mode)
Error indication and warnings after
automatic functionality test
(evaluated and indicated when swit-
ching to active mode)
Indication to enable/disable light
functionality
Indication of the Switch Guard
function being disabled
Indication of the Switch Guard
function being enabled
Indication of unsuccessful light
activation or light forcibly turned off
due to low/critically low batteries
16
MXA51105
Light indication
The indicator light with the battery symbol
flashes slowly and red
(100 ms light/2 s pause).
The indicator light with the battery symbol
flashes quickly and red
(100 ms light/300 ms pause)
The indicator light with the power-on
symbol flashes orange (1 Hz, 500 ms
light/500 ms pause)
The indicator light with the power-on
symbol flashes orange
(5 Hz, 100 ms light/100 ms pause) (Ends
when breathing detected, but 30 times
at most.)
The indicator light with the thermometer
symbol flashes blue
(100 ms light/2 secs pause).
(Lasts until the low temperature is no
longer detected.)
The indicator light with the thermometer
symbol flashes orange
(100 ms light/2 secs pause).
(Lasts until the high temperature is no
longer detected.)
3x red flashes in a circle around the
mechanical switch
(100 ms light/100 ms pause)
Light with the orange power-on symbol
constantly lit. Releasing the button stops
the light.
Critical error detected - Red LEDs flash in a
circle around the mechanical button
Warning detected - no light indication.
3x flashes of the lamp
(300 ms light/300 ms pause)
3x orange flashes of the indicator
light with the power-on symbol
(500 ms light/500 ms pause)
2x orange flashes of the indicator light
with the power-on symbol
(1 sec light/1 sec pause).
5 x flashes of the indicator light with the
battery symbol in red (100 ms light/100
ms pause)
Acoustic indication
No acoustic signal
No acoustic signal
1x beep - warning after 30 secs
from child presence detection, if the
child is still detected (1 sec tone)
3 notification beeps (300 ms
tone/300 ms pause)
(ends when breathing detected, but
3 times at most)
No acoustic signal
No acoustic signal
1x beep - warning (500 ms tone)
No acoustic signal
Acoustic indication according
to the type of error or warning
No acoustic signal
2x notification beeps
(1 sec tone, 1 sec pause)
1x notification beep (3 sec tone)
No acoustic signal